VistOR LED Agency Approvals
Medical Electrical Equipment
With respect to electric shock, fire And mechanical hazards only In accordance with IEC60601-1:2006 / CAN/CSA C22.2 No.601.1-M90 with updates 1 & 2
Classifications:
- Protection against electrical shock (5.1, 5.2). Class I permanently connected,
- Protection against harmful ingress of water (5.3). None.
- Degree of safety in the presence of flammable anesthetics or oxygen (5.5). Not suitable for use in the presence of flammable anesthetics or oxygen.
- Mode of operation (5.6). Continuous
- Surgical luminaries (IEC60601-2-41)
Electromagnetic compatibility for immunity And emissions in accordance with IEC/EN 60601-1-2:2007 ed 3.0 (with 1:2008+A2:2009) and CISPR 11 Class A
Medical Electrical Equipment Particular requirements for the safety of surgical luminaires and luminaires for diagnosis In accordance with IEC-60601-2-41
Intended use: Procedural Medical Lighting for Hospital, Clinic, Minor Surgery, Examination or Diagnosis, within suitable facilities designated for such purposes. This light system is mounted centrally in a room such that access is available all around the operating environment.
User interface: The NUVO lighting systems are minor surgery, procedural and examination lights intended to be used by medical professionals. The functional interfaces are 540° positioning horizontally and +20°/-60° vertically. The lamp head can be rotated approximately 300°. The ON/OFF button also provides for multiple levels of dimming.
Misuses: For any purpose or use at any facility other than stated above. Caution shall be exercised when positioning the light head and avoid contact or collision with the patient, other medical professionals or other lights/equipment.
Table of Contents
Table Caption: This table of contents outlines a user manual for a VISTOR line of lights, covering general information, specifications, installation and assembly for various mounting types, operating instructions, safety and maintenance, and troubleshooting.
Table of Figures
| Figure 1: Ceiling Rod Calculation for Single Ceiling Mount ..................................................... | 11 |
|---|---|
| Figure 2: Ceiling Rod Calculation for Dual Ceiling Mount........................................................ | 12 |
| Figure 3: Recommended Ceiling Structure Construction......................................................... | 14 |
| Figure 4: Mounting Dimensions for Ceiling Mount Assembly................................................... | 15 |
| Figure 5: Arm Cable Termination............................................................................................. | 16 |
| Figure 6: Packing Contents ..................................................................................................... | 17 |
| Figure 7: Mounting Plate Installation ....................................................................................... | 18 |
| Figure 8: Installing the Arm Assembly ..................................................................................... | 19 |
| Figure 9: Finishing the Arm Assembly Installation ................................................................... | 20 |
| Figure 10: Installing the Light Head......................................................................................... | 21 |
| Figure 11: Connecting AC Mains............................................................................................. | 22 |
| Figure 12: Installing the Ceiling Mount Covers ........................................................................ | 22 |
| Figure 13: Wall Bracket Mounting Template............................................................................ | 23 |
| Figure 14: Wall Mount Bracket ................................................................................................ | 23 |
| Figure 15: Extension Arm Mounting on Wall Bracket .............................................................. | 24 |
| Figure 16: Wall mounting with load anchor or stud-anchor...................................................... | 25 |
| Figure 17: Wall mounting with Back-Plate ............................................................................... | 25 |
| Figure 18. Stud Anchor Installation.......................................................................................... | 26 |
| Figure 19: Floor Stand Base Installation.................................................................................. | 27 |
| Figure 20: Floor Stand Electrical Connection .......................................................................... | 28 |
| Figure 21: Installing the Arm Block.......................................................................................... | 29 |
| Figure 22: Articulating Arm Electrical Connection.................................................................... | 30 |
| Figure 23: Adjusting the Articulating Arm................................................................................. | 35 |
| Figure 24: Sterilizable Handle Removal VistOR PRO LED/VistOR PR LED ........................... | 36 |
| Figure 25: Sterilizable Handle Removal VistOR EX LED ........................................................ | 36 |
Table Caption: This table is a list of figures and their corresponding page numbers, likely from a user manual or installation guide for medical lighting equipment. It provides a visual roadmap of the document's content, detailing the illustrations related to ceiling mounts, wall mounts, floor stands, and arm assembly installations.
Section 1: General Information Definition of Terms
I.E.C. International Electrotechnical Commission
ETL Edison Testing Laboratories
Medical Electrical Equipment
Electrical equipment intended to diagnose, treat the patient under medical supervision.
Electrical equipment that transfers light energy to the patient.
Central Illuminance
Illuminance of light head measured at 1 meter from the light emitting area with no obstructions. The value is expressed in Foot-candles or Lux.
Light Field Center
The point of