주요 콘텐츠로 건너뛰기

Access to Housing in the Ulco Integrus PSV

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.
    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. Access to Housing in the Ulco Integrus PSV: 4 단계, 이미지 1/1
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. Access to Housing in the Ulco Integrus PSV: 5 단계, 이미지 1/1
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

결론

To reassemble your device, follow these instructions in reverse order.