Autosampler Reset for Node Triggered Sampling (For Old & New Versions)

필요한 것

  1. Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Switch to manual mode: 1 단계, 이미지 1/2 Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Switch to manual mode: 1 단계, 이미지 2/2
    • To enable manual mode, flip the switch to "Manual Mode"

    • LCD screen will turn on

  2. Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Restart the sampling program: 2 단계, 이미지 1/2 Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Restart the sampling program: 2 단계, 이미지 2/2
    • Press On then press Start Sampling

    • Press Enter

  3. Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Select bottle: 3 단계, 이미지 1/3 Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Select bottle: 3 단계, 이미지 2/3 Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Select bottle: 3 단계, 이미지 3/3
    • Press the number of the bottle you want to sample on

    • Press Enter

    • LCD should now show which bottle it will take a sample in

  4. Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Switch to autonomous mode: 4 단계, 이미지 1/2 Autosampler Reset for Node Triggered Sampling  (For Old & New Versions), Switch to autonomous mode: 4 단계, 이미지 2/2
    • To enable autonomous mode, flip switch to "Node Trigger Mode"

    • LCD should turn off

    • If the autosampler is to be remotely triggered, avoid using the ON/OFF button to power off the sampler.

결론

Now the sampler is ready to be triggered remotely by an open-storm sensor node.