Bear 1000 Main Power Receptacle Replacement

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.

필요한 것

  1. Bear 1000 Main Power Receptacle Replacement, Air and O2 Blender: 1 단계, 이미지 1/1
    • Remove the ribbon cable at the connection on the oxygen blender control board.

  2. Bear 1000 Main Power Receptacle Replacement: 2 단계, 이미지 1/1
    • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.

  3. Bear 1000 Main Power Receptacle Replacement: 3 단계, 이미지 1/1
    • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.

  4. Bear 1000 Main Power Receptacle Replacement, Air and O2 Inlet Manifold: 4 단계, 이미지 1/1
    • Remove the pneumatic and electrical connections labeled.

    • Remove the water trap bowls on the rear of the unit.

  5. Bear 1000 Main Power Receptacle Replacement: 6 단계, 이미지 1/1
    • Remove the four screws and remove the manifold.

    • Remove the two screws holding the power supply to the manifold.

    • Note and mark the orientation of the wires. Remove the two wires going to the fuses at the fuse end and the ground wire going to the main ground stud.

  6. Bear 1000 Main Power Receptacle Replacement: 9 단계, 이미지 1/1
    • Using an allen wrench, remove the mounting screws and nuts from the receptacle.

    • Transfer the three wires form the old receptacle to the new one

결론

To reassemble your device, follow these instructions in reverse order.