Hamilton Medical Ventilator Replacing the LCD

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.

필요한 것

  1. Hamilton Medical Ventilator Replacing the LCD, LCD: 1 단계, 이미지 1/1
    • First open the main enclosure by removing the four hexagonal-drive (Allen) screws at the rear of the enclosure.

  2. Hamilton Medical Ventilator Replacing the LCD: 2 단계, 이미지 1/1
    • Unscrew the control board mounting plate.

  3. Hamilton Medical Ventilator Replacing the LCD: 3 단계, 이미지 1/1
    • “Hinge” back the mounting plate, complete with control board. This exposes the LCD.

  4. Hamilton Medical Ventilator Replacing the LCD: 4 단계, 이미지 1/1
    • Unplug the cable from the GMP adaptor board from the LCD.

  5. Hamilton Medical Ventilator Replacing the LCD: 5 단계, 이미지 1/1
    • Unplug from the dc/ac board, the cable to the interface board.

  6. Hamilton Medical Ventilator Replacing the LCD: 6 단계, 이미지 1/1
    • Unscrew the four hexagonal-drive (Allen) screws securing the two brackets on which the LCD is mounted.

  7. Hamilton Medical Ventilator Replacing the LCD: 7 단계, 이미지 1/1
    • Lift out the complete assembly.

  8. Hamilton Medical Ventilator Replacing the LCD: 8 단계, 이미지 1/1
    • Unplug both of the lights from the dc/ac board.

  9. Hamilton Medical Ventilator Replacing the LCD: 9 단계, 이미지 1/1
    • Remove the LCD from the two brackets, by removing the black hexagonal-drive (Allen) screws on the LCD. NOTE: Do not remove the silver hexagonal-drive (Allen) screws on the SIDE of the LCD.

  10. Hamilton Medical Ventilator Replacing the LCD: 10 단계, 이미지 1/1
    • When fitting a new display, there must be an offset of approximately 2 mm between the plastic sealing strips on the LCD, and the brackets.

  11. Hamilton Medical Ventilator Replacing the LCD: 11 단계, 이미지 1/1
    • If necessary, adjust this gap using the silver screws on the side of the LCD.

    • The surface of the LCD must be clean. Wipe it free of finger marks and dust with an antistatic paper towel.

  12. Hamilton Medical Ventilator Replacing the LCD: 13 단계, 이미지 1/1
    • The cable from the GMP to the LCD must be in the cable grip on the control board mounting plate.

    • When screwing the LCD assembly into GALILEO, you must align it with the glass screen in GALILEO. Check this by looking at the front of GALILEO.

    • Turn on GALILEO, and make sure that the LCD lights up, and is properly aligned.

결론

To reassemble your device, follow these instructions in reverse order.