How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.
    • Adjust the gas flow to 10 L/min.

  1. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: 2 단계, 이미지 1/1
    • Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.

  2. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: 3 단계, 이미지 1/1
    • Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20

  3. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: 4 단계, 이미지 1/1
    • Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.

결론

Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.