Philips Respironics System One REMstar Blower Unit Replacement

필요한 것

  1. Philips Respironics System One REMstar Blower Unit Replacement, Blower Unit: 1 단계, 이미지 1/1
    • Turn the device upside down so that the rubber feet are facing you.

    • Use the TR15 screw head to remove the two 46.5mm length crews located on the back.

    • Gently remove the casing so that the wires are not damaged.

  2. Philips Respironics System One REMstar Blower Unit Replacement: 2 단계, 이미지 1/1
    • Using a Torx TR10 screwdriver, remove the seven 13.5 mm screws that are securing the inner casing.

    • Remove the inner casing.

  3. Philips Respironics System One REMstar Blower Unit Replacement: 3 단계, 이미지 1/1
    • Remove the motor from the bottom half of the inner casing and set it to the side.

  4. Philips Respironics System One REMstar Blower Unit Replacement: 4 단계, 이미지 1/2 Philips Respironics System One REMstar Blower Unit Replacement: 4 단계, 이미지 2/2
    • Remove the circular piece by popping it off to separate the circuit board from the top cover.

    • No tools should be necessary to complete this action.

  5. Philips Respironics System One REMstar Blower Unit Replacement: 5 단계, 이미지 1/2 Philips Respironics System One REMstar Blower Unit Replacement: 5 단계, 이미지 2/2
    • Disconnect the green and red wire connectors from their white connectors to disconnect the blower motor.

결론

To reassemble your device, follow these instructions in reverse order.

다른 2명이 해당 안내서를 완성하였습니다.

댓글 한 개

These steps may be out of sequential order. Step 4 should probably be Step 1 and perhaps some are even missing.