주요 콘텐츠로 건너뛰기

Viasys Avea UIM and Plastic Cover Disassembly

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.

필요한 것

  1. Viasys Avea UIM and Plastic Cover Disassembly, Viasys Avea UIM and Plastic Cover Disassembly: 1 단계, 이미지 1/1
    • Remove the ‘rubber collar’ located at the UIM rear neck, by grasping one of the two rubber tabs at the bottom. Pull firmly in an “arcing” motion.

    • Pull UIM forward and look for the Phillips Screw on the front arm cover. Remove (1) Phillips screw from the front arm cover. Remove the front arm cover.

    • Remove the two mounting screws now visible inside the ‘back arm cover’. Tilt the UIM down and remove the ‘back arm cover’.

  2. Viasys Avea UIM and Plastic Cover Disassembly: 4 단계, 이미지 1/1
    • Remove the (4) 11/32 KEP nuts that secure the ‘plastic top cover’ opening onto the chassis.

    • Remove (2) Phillips screws that secure the ‘plastic top cover’ to the rear chassis.

    • Slide the top ‘plastic top cover’ upward and forward away from the chassis.

    • Remove the (2) Phillips screws and washers that secure the HSSC Cable on the GDE. Unplug the HSSC Cable.

    • Carefully feed the HSSC Cable through the ‘plastic top cover’ opening.

    • While continuously supporting the UIM, remove (4) 3/8 nuts that fasten the UIM support arm onto the pneumatics.

    • Lift UIM off of pneumatics and set aside.

    • Remove plastic cover and set aside.

결론

To reassemble your device, follow these instructions in reverse order.