  1. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement, Power Knob Kit: 1 단계, 이미지 1/1
    • Loosen but do not remove the(4) 8-32 X 3 screws on the outer air intake

    • Pg. 29

  2. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 2 단계, 이미지 1/1
    • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator

    • Pg. 29

  3. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 3 단계, 이미지 1/1
    • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”

    • Pg. 29

  4. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 4 단계, 이미지 1/1
    • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module

    • Pg. 29

  5. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 5 단계, 이미지 1/1
    • Remove the bezel by loosening and removing the (7) 4-40 X ¼ screws that hold the bezel to the front case

    • Pg. 29

  6. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 6 단계, 이미지 1/1
    • On the membrane panel assembly, loosen the 6-32 X ¼ set screw using a 1/16 Ball Hex Driver (located in the power knob) and lift the damaged knob from the membrane panel

    • Pg. 30

  7. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 7 단계, 이미지 1/1
    • Place the replacement knob on the membrane panel switch -align with flat on switch –insert the included 6-32 X ¼ set screw and tighten

    • Pg. 30

  8. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 8 단계, 이미지 1/1
    • Install the new bezel by tightening the included (7) 4-40 X ¼ screws

    • Pg. 30

  9. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 9 단계, 이미지 1/1
    • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position

    • Pg. 30

  10. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 10 단계, 이미지 1/1
    • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10-32 Keps nuts and the (4) 6-32 X 2” screws

    • Pg. 30

  11. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 11 단계, 이미지 1/1
    • Tighten the (4) 8-32 X 3 screws on the outer air intake

    • Pg. 30

  12. Zoll Impact Uni-Vent 731 Power Knob Kit Replacement: 12 단계, 이미지 1/1
    • Perform HiPot Testing then perform Calibration and Functional Test using the RCS

    • Pg. 30

결론

To reassemble your device, follow these instructions in reverse order.