Newport HT70 Pump Assembly Replacement

의료 장비
iFixit과 이 수리 안내서는 의료 기기 제조업체가 보증하지 않으며 제휴 관계가 없습니다.

필요한 것

  1. Newport HT70 Pump Assembly Replacement, Pump Assembly: 1 단계, 이미지 1/1
    • Position the base assembly so that you may access it.

    • Using the #1 Phillips screwdriver, unscrew the four screws (SCR3219P in the photo) and the two screws (SCR3232P) securing the base plate.

  2. Newport HT70 Pump Assembly Replacement: 2 단계, 이미지 1/1
    • The base plate will then come apart from the rest of the device with the pump assembly attached to it, as depicted in the following photo.

    • Inspect pump assembly for any damage or looseness.

  3. Newport HT70 Pump Assembly Replacement: 3 단계, 이미지 1/1
    • If damage is found, replace the pump assembly by removing each tube individually and replacing as needed.

    • You may not need to replace the entire assembly if parts of it are still intact.

결론

To reassemble your device, follow these instructions in reverse order.