Advance Series Bed is a Hospital Bed manufactured by Hill-Rom. Model numbers P1104, P1105, P1115, P1135, P1145, P1155, P1165, P1106, P1116, P1136, P1146.

P & V mode won't work

Error code: 0xA Blower error Dtc Code- 0xA0C5

0xA Blower pressure under limit Dtc Code - 0xA0C8

이 질문에 답하기 저도 같은 문제가 있습니다

좋은 질문입니까?

점수 0

댓글 1개:

olmstead check the blower. Let us know what model bed this is and what model blower it uses. Make sure there is no obstruction between the blower and the valve assembly.

댓글 달기

답변 1개

this service manual is about the MOST complete and well written that I have ever seen. Half the cost of the bed must be in it. However, there is NO reference to any of those codes. There is only LED blink codes. So, do you have something different?

Otherwise, there are 6 pressure sensors and you will have to figure out which one is reading low and go from there.

이 답변이 도움이 되었나요?

점수 0
댓글 달기

답변을 추가하세요