daeg SpO2 Analyzer User and Service Manual
daeg $ \times $ SPO2 Analyzer User and Service Manual
METRON
Metron Offices
USA
1345 Monroe NW, Suite 255A
Grand Rapids, MI 49505
Phone: (+1) 888 863-8766
Fax: (+1) 616 454-3350
E-mail: support@metron.us
FRANCE
30, rue Paul Claude
F-91000 Evry
Phone: (+33) 1 6078 8899
Fax: (+33) 1 6078 6839
E-mail: info@metron.fr
GERMANY
Römerstrasse 22
D-63741 Aschaffenburg
Phone: (+49) 6021 447 9807
Fax: (+49) 6021 447 9808
E-mail: proteges@t-online.de
NORWAY
Vegamot 8
N-7048 Trondheim
Phone: (+47) 7395 4700
Fax: (+47) 7395 4701
E-mail: support@metron.no
Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied, including but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.
Limited Warranty
METRON warrants that the daeg SPO2 Analyzer will substantially conform to published specifications and to the documentation, provided that it is used for the purpose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective analyzer, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the daeg SPO2 Analyzer, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the daeg SPO2 Analyzer. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the daeg SPO2 Analyzer other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.
Table of Contents
Metron Offices.....2
Table of Contents.....3
daeg SpO2 Analyzer Specifications.....5
daeg Quick Start.....7
Unit Setup.....9
Unpacking.....9
Battery Installation.....9
Operation.....11
Power up Screen.....11
Main Simulation Screens.....11
Function Selection Menu.....12
Alarm Limits.....12
Remote Control.....13
Probe Analyzer.....13
Simulation Setup.....14
Print Options.....15
Troubleshooting.....17
Calibration.....19
Schematics.....21
This page was intentionally left blank.
daeg ✗ SpO2 Analyzer Specifications
SPO2 SIMULATION
GENERAL
Factory installed, downloadable and user programmable R-curves, patient conditions and templates.
- Probe signal and LED test x
ACTIVE PROBE SPO2 SIMULATION
O2 saturation range: 30 - 100 %
Simulation resolution: 1% SpO2%
Simulation accuracy:
± 0.5 SpO2% 65-100% SpO2%
$ \pm 1 $ SpO2% 30-64% SpO2%
- Simulation repeatability:
±0.2 SpO2%
Simulation rate: 30 - 300 bpm, res. 5 bpm (sync. to ECG for "C-lock" testing)
Rate accuracy: $ \pm 0.1\% $
Pulse amplitude: 0 - 100% of nominal pleth amplitude.
- Amplitude accuracy: ≥30% amplitude ±1%. 1-29% amplitude ±5%.
Low, medium and high
-
Light repeater bandwidth: DC - 1MHz
-
Optical slew rate: 50 mcd/ $ \mu $s
-
Light repeater dynamic range: 1mcd + 13 dB - 60 dB
-
Receiver optical 3 dB bandwidth: 600 - 1050 nm.
INSTALLED R-CURVES
BCI International
-
Criticare
-
Masimo
-
Nellcor
Novametrix
-
Philips (Agilent)
-
Ohmeda
INSTALLED PATIENT CONDITIONS
-
Normal patient
-
Geriatric patient
Obese patient
Patient with weak pulse
Bradycardia
Tachycardia
ARTIFACTS (ACTIVE PROBE ONLY)
- Motion: 0 - 4Hz
Ambient sunlight: Ambient, Sun
- Interfering light: 50Hz, 60Hz
PROBE LED OPTICAL FREQUENCY
-
Frequency range: 550 - 1050 nm
-
Resolution: 1nm
-
Accuracy: $ \pm2\% $ of range
PROBE TEST
Available with release of Probe Test Adaptor Box and firmware update.
ECG SIMULATION - GENERAL
The data below is related both to factory installed and downloadable waveforms.
Lead configuration:
5 lead RL, RA, LA, LL, V1-6
Low level output amplitude: 0.2-2mV, resolution 100 mV
Low level output impedance: 1000 ohm to RL
High level output amplitude: 0.5V/mV of low level.
High level output impedance: 50 ohm
System amplitude accuracy: $ \pm2\% $ @1mV
System rate accuracy: $ \pm 0.1\% $
System time resolution: 10 µs (100 kHz maximum update rate)
ECG OUTPUT
The data below is related to factory installed waveform types.
NORMAL ECG SINUS
- 30 - 300 BPM, resolution 1 BPM
GENERAL INFORMATION
POWER SUPPLY
- Power sources: 4 standard AA/LR6/MN-1500 cells. 230VAC/115VAC to 9VDC battery eliminator.
Power consumption:
200 mA w/o display backlight.
400 mA w/full backlight
400 mA w/full backlight.
Battery lifetime: 12 h for alkaline cells without backlight.
HOUSING
High impact plastic case.
DIMENSIONS
-
Height: 237 mm/9.2"
-
Width: 122 mm/4.8"
42 mm/1.6"
WEIGHT
With battery: 0.6 kg/1.3 lb.
TEMPERATURE
-
Storage temperature: 0/32 to +50/122 °C/F
-
Operating temperature: +15/59 to +35/95°C/F
HUMIDITY
Operating humidity: 10% - 80% (non-condensing)
CONNECTORS
Graphical LCD, with backlighting.
- 9V DC power inlet
DISPLAY
RS-232/C for PC or printer communication (9-pin D-sub male)
-
1 probe connector (Redel 8-pin).
-
5 low level ECG outputs RA, RL, LA, LL and V1 (AHA color coded safety connector)
1 high level ECG output (standard phono jack).
daeg SpO2 Analyzer
Ordering Information
Order no:
12700: daeg SpO2 Analyzer
Accessories:
17021: Battery Eliminator 240V or
17027: Battery Eliminator 115V
17030: Battery Package. 4-cells
12710: User and Service Manual
Options:
11150: Carrying case
12711: Cable Test Adapter Box
12712: PRO-Soft daeg
12713: PRO-Soft daeg, demo
12714: User Manual PRO-Soft daeg
17024: Universal Banana Adapter