Receiving system error 1018

Receiving System error 1018 to call service. Unplugged and waited 5-10 minutes, machine still doesnโ€™t work.

Thank you,

Camille Kutis

์ด ์งˆ๋ฌธ์— ๋‹ตํ•˜๊ธฐ ์ €๋„ ๊ฐ™์€ ๋ฌธ์ œ๊ฐ€ ์žˆ์Šต๋‹ˆ๋‹ค

์ข‹์€ ์งˆ๋ฌธ์ž…๋‹ˆ๊นŒ?

์ ์ˆ˜ 0
๋Œ“๊ธ€ ๋‹ฌ๊ธฐ

๋‹ต๋ณ€ 1๊ฐœ

๊ฐ€์žฅ ์œ ์šฉํ•œ ๋‹ต๋ณ€

@corkie The following message is displayed on the LCD: "SYSTEM ERR 1018 Call service!"

The error is a Flow sensor fault. The solution is to Replace flow sensor. You can download the service manual from this page ResMed VPAP Auto 25 to work on your CPAP. Ultimately you do want somebody (preferably a BioMed person) look it over after you do your repair.